Evaluation of Dry Textile Electrodes for Long-term Electrocardiographic Monitoring
Key finding
A standard skin-electrode impedance testing framework for dry textile electrodes; impedance alone was found not to be the primary indicator of ECG signal quality.
Abstract
Background: Continuous long-term electrocardiography monitoring has been increasingly recognized for early diagnosis and management of different types of cardiovascular diseases. To find an alternative to Ag/AgCl gel electrodes that are improper for this application scenario, many efforts have been undertaken to develop novel flexible dry textile electrodes integrated into the everyday garments. With significant progresses made to address the potential issues (e.g., low signal to noise ratio, high skin-electrode impedance, motion artifact, and low durability), the lack of standard evaluation procedure hinders the further development of dry electrodes (mainly the design and optimization).
Results: A standard testing procedure and framework for skin-electrode impedance measurement is demonstrated for the development of novel dry textile electrodes. Different representative electrode materials have been screen-printed on textile substrates. To verify the performance of dry textile electrodes, impedance measurements are conducted on an Agar skin model using a universal setup with consistent frequency and pressure. In addition, they are demonstrated for ECG signals acquisition, in comparison to those obtained using conventional gel electrodes.
Conclusions: Dry textile electrodes demonstrated similar impedance when in raised or flat structures. The tested pressure variations had an insignificant impact on electrode impedance. Looking at the effect of impedance on ECG signals, a noticeable effect on ECG signal performance metrics was not observed. Therefore, it is suggested that impedance alone is possibly not the primary indicator of signal quality. As well, the developed methods can also serve as useful guidelines for future textile dry electrode design and testing for practical ECG monitoring applications.
What this proves at MyantX
Long-term wear is where textile ECG has to prove itself, and this KITE-partnered study builds the standard framework for testing it — notably finding that skin-electrode impedance alone is not the primary indicator of signal quality. It grounds the long-term monitoring behind our Textile ECG modality and the connected-garment work in smart apparel. It tells our engineers what to optimize for, and what to stop worrying about.
This research backs
Modalities
Cite this paper
Milad Alizadeh-Meghrazi, Binbin Ying, Alessandra Schlums, Emily Lam, Ladan Eskandarian, Farhana Abbas, Gurjant Sidhu, Amin Mahnam, Bastien Moineau, Milos R. Popovic. “Evaluation of Dry Textile Electrodes for Long-term Electrocardiographic Monitoring.” BioMedical Engineering OnLine, 2021. https://doi.org/10.1186/s12938-021-00905-4
Discuss this work with our team
Talk to the researchers and engineers behind the applied work at MyantX.